Glow Discharge Source The ion source has been designed for sensitivity
and ease of use. A series of plug-in discharge cells enables the analysis of
both pin and flat sample geometries.
Conductors, semiconductors and
superconductors are analyzed directly in the plasma. Inorganic insulating solid
samples are analyzed using the patented secondary cathode system; insulating
powder samples may be compacted with a high-purity conducting binder prior to
analysis. Depth profile analysis is performed using the flat sample geometry.
|
|
|
|
¢º
Source Cryo Cooling All cells are cooled to near liquid nitrogen temperatures using Thermo
Elemental's patented cryo cooling system to minimize molecular interference
species. Cryo-cooling also allows the routine analysis of low melting point
samples such as gallium.
|
|
In the example above, potassium has been
determined with and without cooling. When the system is run "warm," the ArH
molecular species is present at approximately 800 ppm, and the resultant peak
swamps the trace level potassium signal even at high resolution. Cooling
dramatically reduces the ArH polyatomic to 0.5 ppm, enabling the potassium level
to be clearly identified.
|
|
|
|
¢º
Double-Focusing
Mass Analyzer A reverse geometry
Nier Johnson double-focusing mass analyzer first deflects the ion beam by
mass-to-charge ratio around the magnetic sector, focusing at the first focal
point between the magnet and the Electro-Static Analyzer (ESA). The ESA then
deflects and focuses the ion beam by ion energy to the final focal point at the
final slit in front of the detector system. This combination gives the VG9000 a
resolving power of around 8000, easily separating overlapping or interfering
isotopes.
¢º
Faraday and Daly Detector System |
Dual Faraday and
Daly detectors give the VG9000 11 orders linear dynamic range. The Faraday
detector measures the majors within the sample that produce ion currents between
10-13 amps and 10-9 amps. The Daly detector system measures the minor and trace
concentrations. System software automatically calibrates and switches between
the two detectors, providing seamless data acquisition and
analysis.
¢º
Choice of Pumping Systems The VG9000 offers a choice
of two pumping systems:
-
An all
turbo version incorporating a 500 l/s turbo-molecular pump on the source chamber
and two 250 l/s turbo-molecular pumps on the analyzer chamber.
- Alternatively, the system may be fitted with a 1500 l/s cryo-pump on the
source, which provides more efficient pumping of atmospherics and better limits
of detection on carbon, nitrogen and oxygen.
|
|
|
|
|